Look at these two photos. Do you see anything that is out of place in each of ​them?​

dental handpieces comparison
dental tools comparison funny

Hopefully it is obvious. Who would ever consider driving a tractor up Stelvia Pass in Bormio, Italy? This pass is the most famous mountain-driving pass in the world. A Porsche 917 would be way better!

And does it seem appropriate to use a Lola-Drayson Formula racecar to plant a row of vegetables in a vast field? Is that not more appropriate for a John Deere?

The photographs are comical for what they are, but the reality is that there are correlations between the two situations in dentistry. It is common for dentists to use dental handpieces inappropriately for various circumstances. Slowing a “high speed” down to refine crown preparation margins will not create smooth and accurate finish lines. The subsequent restoration is likely to be ill-fitting with open margins. A “high speed” wants to run at high speed! Using “the right tools for the job” and understanding how they work will provide more predictability and better precision to the final restorative results.

Handpieces are the most important “tools” we dentists use in caring for our patients. Removing decay, contouring esthetics and adjusting occlusion are everyday procedures that involve handpiece use. But is a handpiece a handpiece? And can one handpiece “do it all” to meet our needs and requirements of precision and predictability?

​Air-driven vs. electric handpieces for tooth prep

Air-driven handpieces are a mainstay for dentists in America. However, in Europe, 80 percent of dental offices employ the use of electric handpieces to prepare teeth. But there is an increasing trend toward electric handpiece use for U.S. dentists.  There are a few reasons for these increases.

First are the sound and the vibration. As electric handpieces have improved, the high-pitched sound and the vibrational effects have decreased. Even patients comment on the quieter sound and pitch of the electric handpiece. This improvement comes from greater manufacturing precision, improved materials and better gearing mechanisms.

“Bur chatter” is a common phrase applied to air-driven handpiece use. This “chatter” comes from bur “wobble.” This wobble is caused by ball-bearing wear in the air turbine.  The turbines are supported by O-rings that allow for turbine rotation. The more bearing wear, the more wobble. Wobble can also be defined as “bur whipping.” At a high speed, the bur loses concentricity if the bearings and turbine are worn. The more wear, the more whip and the more inaccurate the cutting efficiency and precision. Regular replacement of the turbine and O-rings can rectify the problem, but is it expensive.

On the other hand, electric handpieces use direct gear-to-gear contact to drive the bur. There is a rigid bearing support within the head of the handpiece to provide better concentricity and less friction. This mechanical structure allows for improved trueness and definitive preparation. There is also less heat production and less chance of tooth sensitivity. The tooth preparation is “milled” rather than “chopped!”

But the real difference between air and electric handpieces is speed and torque. Speed refers to revolutions per minute (rpm). Torque is expressed as watts and is an indication of “cutting power.” Air-driven handpieces can have high speeds. These speeds vary between 250,000 and 420,000 rpm. But the torque of these handpieces is low, in the range of 20 watts.

The cutting efficiency of the air-driven handpieces are affected by this lack of torque. Dentists learn to “feather” the handpiece to maintain the high speed. This “on-off-on” feathering takes more time and reduces smoothness of the tooth preparation as well. Applying force while cutting can “stall” an air-driven handpiece. Additionally, pushing a handpiece to the point of stalling creates heat build-up within the tooth. This may result in tooth sensitivity and pulpitis.

Electric handpieces, on the other hand, have lower RPMs but have tremendously higher torque. Electric handpieces have variable speeds from 100 to 200,000 rpm. (Yes, a bur can cut at 100 rpm with an electric!). The torque of electric handpieces is in the 60-watt range (three times that of air-driven handpieces). The bur will not slow down no matter what material, through all the speed ranges. Because the electric handpiece provides consistent torque and power, “feathering” is not necessary. The bur may be applied to the tooth structure and will cut or “mill” smoothly and consistently. 

Because the speed and torque are consistent, minimal heat build-up occurs within the tooth. There is a learning curve to this preparation technique, but it results in tooth preparations that your dental laboratory will notice!

The speed, efficiency and smoothness of the final preparation results outweigh the increased costs for electric versus air-driven handpieces.

There are many different manufacturers of electric handpieces. Weight, head size and ergonomics may be factors for choice. Maintenance of all electric handpieces is similar to air-driven counterparts. Each manufacturer has their individual recommendations for maintenance. Because the head of the handpiece snaps on to the electric motor, there are various attachments that may be utilized. High-speed attachments may be used for crown and veneer preparations and low-speed attachments for decay removal, margin refinement and polishing materials. Additionally, because of the variable speeds of electric motors, endodontic, surgical and implant attachments are all possible.

high speed NSK dental handpiece
low speed NSK dental handpiece
NSK electric motor and cord dentistry

The efficiency of high-speed handpieces far outweighs that of air-driven counterparts. Applied pressure to the tooth structure will not alter the speed of the bur and will result in faster and more efficient tooth preparation. All electric handpieces provide this ability.

A comparative video of electric vs. air-driven power may be viewed on the NSK website. (Go to the video gallery then scroll down to “Ti-max Z vs. competitor air driven handpiece comparison test”). This video shows the power and performance of NSK electric handpieces over air-driven. It is quite impressive!

Low-speed electric handpieces are far superior to “slowing down” or “feathering” both air and electric high speeds. By having the ability to adjust the speed of rotation of the handpiece, greater control of preparation margin smoothness and accuracy is increased.  Generally, the lower the speed, the greater the accuracy. Restoration fit is very noticeable. By using slow speed and refining preparations, the dental laboratory will notice and appreciate the improved impressions and the resultant restoration.

Low speed is also very important for porcelain adjustment and polishing. Micro-fractures are easily introduced into ceramic restorations with occlusal adjustments made with high-speed handpieces (electric or air-driven). These micro-fractures or flaws may create restoration failure. Ceramic polishing is best performed at a known speed of around 5,000 to 9,000 rpms. Speeds greater than this may be catastrophic to the ceramic restoration. Low-speed precision is vital to creating successful, predictable and long-lasting restorations.

Electric handpieces provide precision, accuracy and efficiency to our everyday practice. Having the right tools for the job increase our production and make our outcomes more predictable and provide greater satisfaction for our patients and ourselves.

Technical comparisons: air-driven vs. electric handpieces

​Air-driven handpieces:

  • Bur speed: 400,000 to 420,000 rpm
  • Torque: 14 to 20 watts power
  • Bur slows with applied pressure
  • Air turbine drives
  • Less bur concentricity
  • ​Loud and high-pitched sound
  • Bur "chatter;" "chopping" smooth bur cutting; "milling"
  • Weight: lighter/smaller
  • Less cost

Electric handpieces:

  • Bur speed: 100,000 to 200,000 rpm
  • 60 watts power
  • No bur slowing; maintains torque
  • Direct gear drive
  • ​Greater bur concentricity
  • ​Smoother and quieter sound
  • Motor heavier/larger head?
  • Greater cost (up front)

Jeff Bonk, D.D.S., P.C., Spear Faculty and Contributing Author - http://jeffreybonkdds.com

(Click this link for more dentistry articles by Dr. Jeff Bonk.)



Comments

Commenter's Profile Image Phil P.
November 7th, 2017
Thank you for the technical information Jeff. I've switched from air to electric a couple of years ago and wished I had done it sooner...I may have saved some of my hearing ! Thanks also to all of the Spear faculty and support staff for changing my daily practice life for the better! See each of you again soon.
Commenter's Profile Image Trevor H.
November 13th, 2017
After the Restorative Dentistry course, we got the electric HPs for our office. Happy I did, my preps are so much better, the lab actually thanked me. Was wondering if you have issues with the water spray when using the christmas tree depth cutting burs? My water spray seems to hit the widest diamond at the top and doesn't get to the thinner part at the bottom. -Trevor